WJPR Citation
|
| All | Since 2020 | |
| Citation | 8502 | 4519 |
| h-index | 30 | 23 |
| i10-index | 227 | 96 |
A DEVELOPMENT AND VALIDATION OF AN RP-HPLC METHOD FOR ESTIMATION OF FENOPROFEN IN BULK DRUG
Syed Rihana*, Dr. A. Suneetha, G. Sandhyarani, G. Sumanjali, G. Jyothi, G. Chaitanya
Abstract Rapid and sensitive method was developed and validated for the estimation of fenoprofen calcium in bulk drugs. Liquid-liquid extraction with methanol and acetonitrile. The accuracy and reproducibility of the method were within acceptable. Linearity ranges from 10 to 80 microgram/ml .the run time of the sample is 10 minutes.selected wavelength of the drug is 270nm. Operating temperature at 25degree celcius (ambient) by using column –enable c18 (4.6×250mm, 5 micrometers). Limit of the developed rp-hplc method for the estimation of fenoprofen, carried out on with enable c18 (250nm×4.6, 5μm) column. Methanol and acetonitrile in the ratio (80:20v/v) used as mobile phase and flow rate 1.5 ml/min. The detection was carried at 270nm and ambient column temperature was maintained. Keywords: linearity, accuracy, precision, limit of detection, limit of quantification. [Full Text Article] [Download Certificate] |